4.7 Article

Automated quality control of brain MR images

期刊

JOURNAL OF MAGNETIC RESONANCE IMAGING
卷 28, 期 2, 页码 308-319

出版社

WILEY
DOI: 10.1002/jmri.21434

关键词

automated; quality control; magnetic resonance imaging; computer-assisted image analysis; quantitative evaluation; brain

向作者/读者索取更多资源

Purpose: To present a novel fully automated method for assessing the quality of magnetic resonance imaging (MRI) data acquired in a clinical trials environment. Materials and Methods: This work was performed in the context of clinical trials for multiple sclerosis. Quality control (QC) procedures included were: (i) patient brain identity verification, (ii) alphanumeric parameter matching, (iii) signal-to-noise ratio estimation, (iv) gadolinium-enhancement verification, and (v) detection of ghosting due to head motion. Each QC procedure produces a quantitative measurement which is compared against an acceptance threshold that was determined based on receiver operating characteristic analysis of traditional manual and visual QC performed by trained experts. Results: The automated QC results have high sensitivity and specificity when compared with the visual QC. Conclusion: Our automated objective QC procedure can replace many manual subjective procedures to provide increased data throughput while reducing reader variability.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据