期刊
JOURNAL OF MAGNETIC RESONANCE
卷 196, 期 1, 页码 84-87出版社
ACADEMIC PRESS INC ELSEVIER SCIENCE
DOI: 10.1016/j.jmr.2008.09.019
关键词
Solid state NMR; Magnetic resonance; Magic-angle spinning; Temperature calibration
资金
- National Institute of Diabetes and Digestive and Kidney Diseases (NIDDK)
- NIH
- NATIONAL INSTITUTE OF DIABETES AND DIGESTIVE AND KIDNEY DISEASES [ZIADK029029] Funding Source: NIH RePORTER
Accurate determination of sample temperatures in solid state nuclear magnetic resonance (NMR) with magic-angle spinning (MAS) can be problematic, particularly because frictional heating and heating by radio-frequency irradiation can make the internal sample temperature significantly different from the temperature outside the MAS rotor. This paper demonstrates the use of Br-79 chemical shifts and spin-lattice relaxation rates in KBr powder as temperature-dependent parameters for the determination of internal sample temperatures. Advantages of this method include high signal-to-noise, proximity of the Br-79 NMR frequency to that of C-13, applicability from 20 K to 320 K or higher, and simultaneity with adjustment of the MAS axis direction. We show that spin-lattice relaxation in KBr is driven by a quadrupolar mechanism. We demonstrate a simple approach to including KBr powder in hydrated samples. such as biological membrane samples, hydrated amyloid fibrils, and hydrated microcrystalline proteins, that allows direct assessment of the effects of frictional and radio-frequency heating under experimentally relevant conditions. Published by Elsevier Inc.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐