4.7 Article

Zincon-modified activated carbon for solid-phase extraction and preconcentration of trace lead and chromium from environmental samples

期刊

JOURNAL OF HAZARDOUS MATERIALS
卷 166, 期 1, 页码 133-137

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.jhazmat.2008.11.006

关键词

Lead; Chromium; Modified activated carbon; Solid-phase extraction; ICP-OES

向作者/读者索取更多资源

A new method that utilizes zincon-modified activated carbon (AC-ZCN) as a solid-phase extractant has been developed for simultaneous preconcentration of trace Cr(III) and Pb(II) prior to the measurement by inductively coupled plasma optical emission spectrometry (ICP-OES). The separation/preconcentration conditions of analytes were investigated, including effects of pH, the shaking time, the sample flow rate and volume, the elution condition and the interfering ions. At pH 4, the maximum adsorption capacity of Cr(III) and Pb(II) onto the AC-ZCN were 17.9 and 26.7 mg g(-1), respectively. The adsorbed metal ions were quantitatively eluted by 1 mL of 0.1 mol L-1 HCl. Common coexisting ions did not interfere with the separation. According to the definition of IUPAC, the detection limits (3 sigma) of this method for Cr(III) and Pb(II) were 0.91 and 0.65 ng mL(-1), respectively. The relative standard deviation under optimum condition is less than 3.5% (n = 8). The method has been applied for the determination of Cr(III) and Pb(II) in biological materials and water samples with satisfactory results. (C) 2008 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据