期刊
JOURNAL OF HAZARDOUS MATERIALS
卷 161, 期 2-3, 页码 1391-1398出版社
ELSEVIER SCIENCE BV
DOI: 10.1016/j.jhazmat.2008.04.103
关键词
Heavy metal; Ultrasound energy; Environmental sample; Biological sample; Certified reference material
In this study, a sample preparation method based on ultrasonic assisted acid digestion (UAD) has been evaluated for total heavy metals (Cd, Cr, Ni and Ph) determination in different environmental (soil, sediment and sewage sludge), and biological (fish muscles, vegetables and grains) samples, using electrothermal atomic absorption spectrometry (ETAAS). The investigated parameters influencing UAD such as presonication time, sonication time, temperature of ultrasonic bath, and different acid mixtures were fully optimized, whereas power was maintained constant at 100% of nominal power of ultrasonic bath. Six different sets of above parameters were applied on six certified reference materials (CRMs) having different matrices. The accuracy of the method was also tested by comparing the results with those obtained from conventional hot plate assisted acid digestion method on same CRMs. Analytical results for HMs by both methods showed no significant difference at 95% confidence limit (p < 0.05). Recoveries of HMs ranging from 96.2% to 102% and 96.3% to 98.6% were obtained from biological and environmental samples, respectively. The average relative standard deviation of UAD method varied between 3.5% and 8.2%, depending on the analyte. (c) 2008 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐