Repeatability and Reproducibility of Goldmann Applanation, Dynamic Contour, and Ocular Response Analyzer Tonometry

期刊

JOURNAL OF GLAUCOMA
卷 22, 期 2, 页码 127-132

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/IJG.0b013e3182254ba3

关键词

intraocular pressure; repeatability; reproducibility; Goldmann applanation tonometry; dynamic contour tonometry; ocular response analyzer; waveform score

资金

  1. NIH NEI [EY08208]

向作者/读者索取更多资源

Purpose: To evaluate the repeatability and interoperator reproducibility of the Pascal dynamic contour tonometry (DCT), ocular response analyzer (ORA), and Goldmann applanation tonometer (GAT) in a single population of normal individuals. Methods: The study included 52 eyes from 26 normal individuals. One operator measured the intraocular pressure (IOP) with each tonometer 3 times while 2 additional operators each measured the IOP with each tonometer once. Repeatability and reproducibility were assessed by the coefficient of variation (CV) and intraclass correlation coefficient (ICC). Agreement among tonometers was also assessed using Bland-Altman plots. Results: The mean age of included participants was 31.5 +/- 8.8 years and 15 (58%) were female individuals. In general, both intra-operator repeatability and interoperator reproducibility were significantly higher for DCT compared with the other tonometers. Intraoperator DCT (CV = 3.7, ICC = 0.89), GAT (CV = 9.7, ICC = 0.79), IOPg (CV = 7.0, ICC = 0.79), and IOPcc (CV = 9.8, ICC = 0.57). Interoperator DCT (CV = 6.1, ICC = 0.73), GAT (CV = 9.0, ICC = 0.82), and IOPg (CV = 10.8, ICC = 0.63), IOPcc (CV = 11.7, ICC = 0.49). Conclusion: Overall, DCT was significantly more repeatable and reproducible than GAT, IOPg and IOPcc. The better reproducibility of the DCT may result in more precise measurements for monitoring IOP changes over time compared to GAT and ORA.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据