Clinical comparison of the Icare tonometer and Goldmann applanation tonometry

期刊

JOURNAL OF GLAUCOMA
卷 17, 期 1, 页码 43-47

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/IJG.0b013e318133fb32

关键词

intraocular pressure; impact tonometer; rebound tonometer; Icare tonometer; Goldmann applanation tonometer

向作者/读者索取更多资源

Purpose: To compare a new method of intraocular pressure (IOP) measurement, using the Icare tonometer, with Goldmann applanation tonometry (GAT). Patients and Methods:Two observers obtained IOP readings in 292 eyes (143 right and 149 left) of 153 subjects, using the]care without topical anesthetic. A GAT reading was subsequently obtained by a consultant ophthalmologist, without the knowledge of the Icare readings. Central corneal thickness (CCT) was obtained on all eyes with ultrasound pachymetry. Patient comfort after IOP measurement was assessed in a consecutive subset of patients. Results: The intraclass correlation coefficient between the 2 modalities of IOP measurement was r = 0.95 for the right and r = 0.93 for the left eye. The mean difference (Icare - GAT) between the IOP measured by the 2 methods was 0.4mm Hg in the right eye (SD 3.0, 9.5% confidence interval - 5.5 to 6.3), and 0.8 mm Hg in the left eye (SD 3.0, confidence interval - 4.7 to 6.2). GAT measurements did not vary with CCT [correlation coefficient = 0.09 (P = 0.25) right and 014 (P = 0.09) left eyes]. However, IOP measured with Icare tonometry increased with increasing CCT [correlation coefficient = 0.16 (P = 0.05) right and 0.21 (P = 0.01) left eyes]. For every 100-mu m increase in CCT, the difference (Icare - GAT) increased by I mm Hg. Of the 38 consecutive patients surveyed, 28 (73.7%) rated the Icare more comfortable than GAT, with only 2 (5.3%) rating it less comfortable (P < 0.001). Conclusions: There is good correlation between the 2 methods of IOP measurement, even at extremes of IOP. The Icare instrument was easy to use and recorded rapid and consistent readings with minimal training. It seems to be more comfortable than GAT and obviates the need for topical anesthesia.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据