4.4 Article

Comparison of In Vivo Optical Systems for Bioluminescence and Fluorescence Imaging

期刊

JOURNAL OF FLUORESCENCE
卷 23, 期 5, 页码 909-920

出版社

SPRINGER/PLENUM PUBLISHERS
DOI: 10.1007/s10895-013-1215-9

关键词

Optical imaging; In vivo; Bioluminescence; Fluorescence; Spectral unmixing

资金

  1. EU
  2. FWO [G.0621.10]
  3. Bijzonder Onderzoeksfonds (BOF) from the Ghent University

向作者/读者索取更多资源

In vivo optical imaging has become a popular tool in animal laboratories. Currently, many in vivo optical imaging systems are available on the market, which often makes it difficult for research groups to decide which system fits their needs best. In this work we compared different commercially available systems, which can measure both bioluminescent and fluorescent light. The systems were tested for their bioluminescent and fluorescent sensitivity both in vitro and in vivo. The IVIS Lumina II was found to be most sensitive for bioluminescence imaging, with the Photon Imager a close second. Contrary, the Kodak system was, in vitro, the most sensitive system for fluorescence imaging. In vivo, the fluorescence sensitivity of the systems was similar. Finally, we examined the added value of spectral unmixing algorithms for in vivo optical imaging and demonstrated that spectral unmixing resulted in at least a doubling of the in vivo sensitivity. Additionally, spectral unmixing also enabled separate imaging of dyes with overlapping spectra which were, without spectral unmixing, not distinguishable.