  1. National Natural Science Foundation of China (NSFC) [40771180, 40688001, 40821003]
  2. Guangdong Provincial Natural Science Foundation [8251064004000001]

向作者/读者索取更多资源

A multiresidue analytical method was developed for the determination of 9 endocrine disrupting chemicals (EDCs) and 19 pharmaceuticals and personal care products (PPCPs) including acidic and neutral pharmaceuticals in water and soil samples using rapid resolution liquid chromatography-tandem mass spectrometry (RRLC-MS/MS). Solid phase extraction (SPE), and ultrasonic extraction combined with silica gel purification were applied as pretreatment methods for water and soil samples, respectively. The extracts of the EDCs and PPCPs in water and soil samples were then analyzed by RRLC-MS/MS in electrospray ionization (ESI) mode in three independent runs. The chromatographic mobile phases consisted of Milli-Q water and acetonitrile for EDCs and neutral pharmaceuticals, and Milli-Q water containing 0.01% acetic acid (v/v) and acetonitrile: methanol (1: 1, v/v) for acidic pharmaceuticals at a flow rate of 0.3 mL/min. Most of the target compounds exhibited signal suppression due to matrix effects. Measures taken to reduce matrix effects included use of isotope-labeled internal standards, and application of matrix-match calibration curves in the RRLC-MS/MS analyses. The limits of quantitation ranged between 0.15 and 14.08 ng/L for water samples and between 0.06 and 10.64 ng/g for solid samples. The recoveries for the target analytes ranged from 62 to 208% in water samples and 43 to 177% in solid samples, with majority of the target compounds having recoveries ranging between 70-120%. Precision, expressed as the relative standard deviation (RSD), was obtained less than 7.6 and 20.5% for repeatability and reproducibility, respectively. The established method was successfully applied to the water and soil samples from four irrigated plots in Guangzhou. Six compounds namely bisphenol-A, 4-nonylphenol, triclosan, triclocarban, salicylic acid and clofibric acid were detected in the soils.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据