期刊
JOURNAL OF ENDODONTICS
卷 37, 期 9, 页码 1293-1295出版社
ELSEVIER SCIENCE INC
DOI: 10.1016/j.joen.2011.05.023
关键词
Endodontic instruments; flexibility; nickel-titanium; thermal treatments
Introduction: The aim of the present study was to evaluate the bending properties of Hyflex instruments, which exhibit a lower percent in weight of nickel (52 Ni %wt) and compare them with other commercially available nickel-titanium (NiTi) rotary instruments. Methods: Ten instruments with tip size 25, 0.06 taper of each of the following NiTi rotary instrumentation techniques were selected for the study: Hyflex, Endo Sequence, Pro File, Hero, and Flexmasters. All instruments from each group were tested for stiffness by comparing their bending moment when they attained a 45-degree bend. Experimental procedures strictly followed testing methodology described in ISO 3630-1. All data were recorded and subjected to statistical evaluation by using analysis of variance test. Statistical significance was set at P < .05). Results: Statistical analysis of the data revealed that Hyflex files were found to be the most flexible instruments, with a significant difference (P < .05) in comparison with the other instruments. Among the other files, a significant difference has been reported for Endo Sequence instruments compared with Pro File, Hero, and Flex Master (P < .05), whereas no significant differences have been reported among those 3 files (P > .05). Conclusions: Results of the present study have illustrated an increased flexibility of the new NiTi alloy over conventional NiTi alloy, and they highlight the potential of the new manufacturing process. (J Endod 2011;37:1293-1295)
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。