4.5 Article

Detection of Vertical Root Fractures of Different Thicknesses in Endodontically Enlarged Teeth by Cone Beam Computed Tomography versus Digital Radiography

期刊

JOURNAL OF ENDODONTICS
卷 36, 期 7, 页码 1245-1249

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.joen.2010.03.021

关键词

Cone-beam computed tomography; digital radiography; thickness; vertical root fracture

向作者/读者索取更多资源

Introduction: This study was performed to compare the accuracy of cone beam computed tomography (CBCT) scans and digital radiography (DR) for detecting simulated vertical root fractures (VRFs) with different thicknesses in extracted human teeth. Methods: Eighty teeth were endodontically prepared and divided into 4 groups, 3 experimental and 1 control. The teeth in experimental groups were artificially fractured and fixed together with different thicknesses of 0.2 mm (0.2-mm VRF group), 0.4 mm (0.4-mm VRF group), and smaller than 0.2 mm (crack group). Teeth in control group were kept intact. Three observers evaluated the DR and CBCT images in terms of accuracy for VRF detection. Results: No significant differences in detection rate were noted among the observers (P > .5). The overall accuracy for detecting VRFs was significantly higher for CBCT scans compared with DR (P < .001). CBCT scans were more accurate in detecting 0.2-mm VRFs (70%) and 0.4-mm VRFs (90%) as compared with DR (43.3% and 60%, respectively). Conclusions: The results of this study showed that CBCT scans are effective for detecting VRFs of different thicknesses. (J Endod 2010;36:1245-1249)

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据