4.5 Article

Influence of Voxel Size in the Diagnostic Ability of Cone Beam Tomography to Evaluate Simulated External Root Resorption

期刊

JOURNAL OF ENDODONTICS
卷 35, 期 2, 页码 233-235

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.joen.2008.11.005

关键词

Cone beam Computed tomography; diagnosis; external root resorption

向作者/读者索取更多资源

This study assessed in vitro the diagnostic ability of cone beam computed tomography (CBCT) scans with different voxel resolutions in the detection of simulated external root resorption (ERR). For that purpose, 59 teeth were viewed through iCAT tomography (Imaging Sciences International, Inc, Hatfield, PA) following three protocols in which the variation was the voxel resolution (0.4, 0.3, and 0.2mm). A calibrated examiner blinded to the protocol assessed the images through the i-CAT View software (Imaging Sciences International, Inc). The chi-square statistical analysis did not show associations between voxel resolution, section plane, size of cavity, and radicular third. Sensitivity and specificity values were similar. However, likelihood ratio values of 6.4 for a 0.4-mm voxel, 16 for a 0.3-mm voxel, and 12 for a 0.2-mm voxel were found. It was concluded that CBCT is it reliable method for the investigation of simulated ERR, and a 0.3-mm voxel appeared to be the best protocol, associating good diagnostic performance with lower X-ray exposure. (J Endod 2009;35:233-235)

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据