4.1 Article Proceedings Paper

Comparison of 5 methods of QT interval measurements on electrocardiograms from a thorough QT/QTc study: effect on assay sensitivity and categorical outliers

期刊

JOURNAL OF ELECTROCARDIOLOGY
卷 44, 期 2, 页码 96-104

出版社

CHURCHILL LIVINGSTONE INC MEDICAL PUBLISHERS
DOI: 10.1016/j.jelectrocard.2010.11.010

关键词

Long QT syndrome; Electrocardiography; Arrhythmias; Sudden cardiac death; QT prolongation; U wave; Cardiac repolarization

向作者/读者索取更多资源

Introduction: We studied moxifloxacin-induced QT prolongation and proportion of categorical QTc outliers when 5 methods of QT measurement were used to analyze electrocardiograms (ECGs) from a thorough QT study. Methods: QT interval was measured by the threshold, tangent, superimposed median beat, automated global median beat, and longest QT methods in a central ECG laboratory in 2730 digital ECGs from 39 subjects during placebo and moxifloxacin treatment. Results: All 5 methods were able to demonstrate statistically significant moxifloxacin-induced QTcF prolongation. However, lower bound of 95% 1-sided confidence interval of QTcF prolongation did not exceed 5 milliseconds with the longest QT method. More QTcF outliers were observed with the longest QT and tangent methods, whereas the other 3 methods were comparable. QTcF values greater than 500 milliseconds were observed only with moxifloxacin by the tangent method, and with moxifloxacin and placebo by the longest QT method. Conclusion: The method of QT measurement must be considered when interpreting individual thorough QT/QTc studies. (C) 2011 Elsevier Inc. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据