期刊
JOURNAL OF CRANIO-MAXILLOFACIAL SURGERY
卷 36, 期 8, 页码 443-449出版社
CHURCHILL LIVINGSTONE
DOI: 10.1016/j.jcms.2008.04.003
关键词
craniofacial; precision; rapid prototyping
资金
- CNPq
- CenPRA
Background: Selective laser sintering (SLS) and three-dimensional printing (3DP (TM)) are rapid prototyping (RP) techniques to fabricate prototypes from biomedical images. To be used in maxillofacial surgery, these models must accurately reproduce the craniofacial skeleton. Purpose: To analyze the capacity of SLS and 3DP (TM) models to reproduce craniomaxillary anatomy and their dimensional error. Material: Dry skull, helical computed-tomography images, SLS and 3DP (TM) prototypes, and electronic calliper. Methods: Tomographic images of a dry skull were manipulated with the In Vesalius biomedical software. Prototypes were fabricated using SLS, and 3DP (TM) techniques. Ten linear measurements were made on the models and compared with corresponding dry skull measurements (criterion standard) carried out with an electronic calliper. Results: We observed a dimensional error of 2.10 and 2.67% for SLS and 3DP (TM) models, respectively. The models satisfactorily reproduced anatomic details, except for thin bones, small foramina and acute bone projections. The SLS prototypes showed greater dimensional precision and reproduced craniomaxillary anatomy more accurately than the 3DP (TM), models. Conclusion: Both SLS and 3DP (TM) models provided acceptable precision and may be useful aids in most maxillofacial surgeries. (C) 2008 European Association for Cranio-Maxillofacial Surgery
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐