Comparison of Hybrid and Pure Iterative Reconstruction Techniques With Conventional Filtered Back Projection: Dose Reduction Potential in the Abdomen

期刊

JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY
卷 36, 期 3, 页码 347-353

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/RCT.0b013e31824e639e

关键词

CT image reconstruction; fully iterative reconstruction; radiation dose reduction

资金

  1. GE Healthcare
  2. Siemens
  3. Philips

向作者/读者索取更多资源

Purpose: Assess the effect of filtered back projection (FBP) and hybrid (adaptive statistical iterative reconstruction [ASIR]) and pure (model-based iterative reconstruction [MBIR]) iterative reconstructions on abdominal computed tomography (CT) acquired with 75% radiation dose reduction. Materials and Methods: In an institutional review board-approved prospective study, 10 patients (mean [standard deviation] age, 60 (8) years; 4 men and 6 women) gave informed consent for acquisition of additional abdominal images on 64-slice multidetector-row CT (GE 750HD, GE Healthcare). Scanning was repeated over a 10-cm scan length at 200 and 50 milliampere second (mA s), with remaining parameters held constant at 120 kilovolt (peak), 0.984:1 pitch, and standard reconstruction kernel. Projection data were deidentified, exported, and reconstructed to obtain 4 data sets (200-mA s FBP, 50-mA s FBP, 50-mA s ASIR, 50-mA s MBIR), which were evaluated by 2 abdominal radiologists for lesions and subjective image quality. Objective noise and noise spectral density were measured for each image series. Results: Among the 10 patients, the maximum weight recorded was 123 kg, with maximum transverse diameter measured as 43.7 cm. Lesion conspicuity at 50-mA s MBIR was better than on 50-mA s FBP and ASIR images (P < 0.01). Image noise was rated as suboptimal on low-dose FBP and ASIR but deemed acceptable in MBIR images. Objective noise with 50-mA s MBIR was 2 to 3 folds lower compared to 50-mA s ASIR, 50-mA s FBP, and 200-mA s FBP (P < 0.0001).Noise spectral density analyses demonstrated that ASIR retains the noise spectrum signature of FBP, whereas MBIR has much lower noise with a more regularized noise spectrum pattern. Conclusion: Model-based iterative reconstruction renders acceptable image quality and diagnostic confidence in 50-mA s abdominal CT images, whereas FBP and ASIR images are associated with suboptimal image quality at this radiation dose level.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据