Multi-centre reproducibility of diffusion MRI parameters for clinical sequences in the brain

期刊

NMR IN BIOMEDICINE
卷 28, 期 4, 页码 468-485

出版社

WILEY
DOI: 10.1002/nbm.3269

关键词

diffusion; MRI; reproducibility; brain; multi-centre

资金

  1. Cancer Research UK [C7809/A10342]
  2. MRC [G1002276, G0300117] Funding Source: UKRI
  3. Cancer Research UK [16464, 10342] Funding Source: researchfish
  4. Medical Research Council [G0300117, G1002276] Funding Source: researchfish
  5. National Institute for Health Research [NF-SI-0512-10162, NIHR-RP-02-12-019] Funding Source: researchfish

向作者/读者索取更多资源

The purpose of this work was to assess the reproducibility of diffusion imaging, and in particular the apparent diffusion coefficient (ADC), intra-voxel incoherent motion (IVIM) parameters and diffusion tensor imaging (DTI) parameters, across multiple centres using clinically available protocols with limited harmonization between sequences. An ice-water phantom and nine healthy volunteers were scanned across fives centres on eight scanners (four Siemens 1.5T, four Philips 3T). The mean ADC, IVIM parameters (diffusion coefficient D and perfusion fraction f) and DTI parameters (mean diffusivity MD and fractional anisotropy FA), were measured in grey matter, white matter and specific brain sub-regions. A mixed effect model was used to measure the intra-and inter-scanner coefficient of variation (CV) for each of the five parameters. ADC, D, MD and FA had a good intra-and inter-scanner reproducibility in both grey and white matter, with a CV ranging between 1% and 7.4%; mean 2.6%. Other brain regions also showed high levels of reproducibility except for small structures such as the choroid plexus. The IVIM parameter f had a higher intra-scanner CV of 8.4% and inter-scanner CV of 24.8%. No major difference in the inter-scanner CV for ADC, D, MD and FA was observed when analysing the 1.5T and 3T scanners separately. ADC, D, MD and FA all showed good intra-scanner reproducibility, with the inter-scanner reproducibility being comparable or faring slightly worse, suggesting that using data from multiple scanners does not have an adverse effect compared with using data from the same scanner. The IVIM parameter f had a poorer inter-scanner CV when scanners of different field strengths were combined, and the parameter was also affected by the scan acquisition resolution. This study shows that the majority of diffusion MRI derived parameters are robust across 1.5T and 3T scanners and suitable for use in multi-centre clinical studies and trials. (C) 2015 The Authors NMR in Biomedicine Published by John Wiley & Sons Ltd.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据