4.7 Article

Optimization of Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry Analysis for Bacterial Identification

期刊

JOURNAL OF CLINICAL MICROBIOLOGY
卷 50, 期 12, 页码 3845-3852

出版社

AMER SOC MICROBIOLOGY
DOI: 10.1128/JCM.00626-12

关键词

-

向作者/读者索取更多资源

Matrix-assisted laser desorption ionization-time of flight mass spectrometry (MALDI-TOF MS) is a relatively new addition to the clinical microbiology laboratory. The performance of the MALDI Biotyper system (Bruker Daltonics) was compared to those of phenotypic and genotypic identification methods for 690 routine and referred clinical isolates representing 102 genera and 225 unique species. We systematically compared direct-smear and extraction methods on a taxonomically diverse collection of isolates. The optimal score thresholds for bacterial identification were determined, and an approach to address multiple divergent results above these thresholds was evaluated. Analysis of identification scores revealed optimal species- and genus-level identification thresholds of 1.9 and 1.7, with 91.9% and 97.0% of isolates correctly identified to species and genus levels, respectively. Not surprisingly, routinely encountered isolates showed higher concordance than did uncommon isolates. The extraction method yielded higher scores than the direct-smear method for 78.3% of isolates. Incorrect species were reported in the top 10 results for 19.4% of isolates, and although there was no obvious cutoff to eliminate all of these ambiguities, a 10% score differential between the top match and additional species may be useful to limit the need for additional testing to reach single-species-level identifications.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据