4.7 Article

Comparison of the Vitek 2 Antifungal Susceptibility System with the Clinical and Laboratory Standards Institute (CLSI) and European Committee on Antimicrobial Susceptibility Testing (EUCAST) Broth Microdilution Reference Methods and with the Sensititre YeastOne and Etest Techniques for In Vitro Detection of Antifungal Resistance in Yeast Isolates

期刊

JOURNAL OF CLINICAL MICROBIOLOGY
卷 48, 期 5, 页码 1782-1786

出版社

AMER SOC MICROBIOLOGY
DOI: 10.1128/JCM.02316-09

关键词

-

资金

  1. Fondo de Investigaciones Sanitarias [FI05/00856]
  2. Red Espanola de Investigacion de Patologia Infecciosa (REIPI) [RD06/0008]
  3. Spanish Network for Research in Infectious Diseases [REIPI RD06/0008]
  4. Astellas Pharma
  5. bioMerieux
  6. Gilead Sciences
  7. Merck Sharp and Dohme
  8. Pfizer
  9. Schering Plough
  10. Soria Melguizo S.A.
  11. European Union
  12. ALBAN program
  13. Spanish Agency for International Cooperation
  14. Spanish Ministry of Culture and Education
  15. Spanish Health Research Fund
  16. Instituto de Salud Carlos III
  17. Ramon Areces Foundation
  18. Mutua Madrilena Foundation

向作者/读者索取更多资源

The commercial technique Vitek 2 system for antifungal susceptibility testing of yeast species was evaluated. A collection of 154 clinical yeast isolates, including amphotericin B-and azole-resistant organisms, was tested. Results were compared with those obtained by the reference procedures of both the CLSI and the European Committee on Antimicrobial Susceptibility Testing (EUCAST). Two other commercial techniques approved for clinical use, the Etest and the Sensititre YeastOne, were included in the comparative exercise as well. The average essential agreement (EA) between the Vitek 2 system and the reference procedures was >95%, comparable with the average EAs observed between the reference procedures and the Sensititre YeastOne and Etest. The EA values were >97% for Candida spp. and stood at 92% for Cryptococcus neoformans. Intraclass correlation coefficients (ICC) between the commercial techniques and the reference procedures were statistically significant (P < 0.01). Percentages of very major errors were 2.6% between Vitek 2 and the EUCAST technique and 1.6% between Vitek 2 and the CLSI technique. The Vitek 2 MIC results were available after 14 to 18 h of incubation for all Candida spp. (average time to reading, 15.5 h). The Vitek 2 system was shown to be a reliable technique to determine antifungal susceptibility testing of yeast species and a more rapid and easier alternative for clinical laboratories than the procedures developed by either the CLSI or EUCAST.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据