Evaluation of the Diagnostic Performance of the Xpert Clostridium difficile Assay and Its Comparison With the Toxin A/B Enzyme-Linked Fluorescent Assay and In-House Real-Time PCR Assay Used for the Detection of Toxigenic C. difficile

期刊

JOURNAL OF CLINICAL LABORATORY ANALYSIS
卷 28, 期 2, 页码 124-129

出版社

WILEY
DOI: 10.1002/jcla.21655

关键词

VIDAS; Clostridium difficile; toxin; Xpert

资金

  1. Inje University research

向作者/读者索取更多资源

Background Clostridium difficile genes or toxin can be detected using several laboratory techniques. In this study, we compared the performance of the Xpert C. difficile assay with that of a toxin A/B enzyme-linked fluorescent immunoassay (ELFA) and an in-house real-time PCR assay for the tcdB gene. Methods From April 2011 through January 2012, 138 soft or liquid stool samples from 138 adult patients at Paik Hospital were tested using the toxin A/B ELFA, in-house real-time PCR assay, and Xpert C. difficile assay to detect toxigenic C. difficile. Specimens were considered true positives if results were positive in both the in-house real-time PCR for tcdB gene and Xpert C. difficile assays. Results Sensitivity of the toxin A/B ELFA, in-house tcdB gene real-time PCR, and Xpert C. difficile assay were 67.6%, 97.3%, and 100.0%, respectively. The specificity of the in-house tcdB gene real-time PCR assay was 100%, while the specificity was 98.0% for the other two methods. The turnaround time (TAT) was 50 min for the Xpert C. difficile assay, 75 min for the toxin A/B ELFA, and 160 min for the in-house real-time PCR assay. Conclusion The Xpert C. difficile assay and the in-house real-time PCR assay had higher sensitivity than the toxin A/B ELFA; however, the specificities of the three assays were similar. Considering its rapid TAT and high sensitivity, use of the Xpert C. difficile assay is highly recommended for rapid and accurate diagnosis of C. difficile infection.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据