Biomonitoring method for bisphenol A in human urine by ultra-high-performance liquid chromatography-tandem mass spectrometry

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.jchromb.2014.01.039

关键词

Bisphenol A; Liquid chromatography; Tandem mass spectrometry; Human urine

资金

  1. National Institute of Environmental Health Sciences - National Institutes of Health [R01ES020488]

向作者/读者索取更多资源

An ultra-high-performance liquid chromatography-tandem mass spectrometry method for the measurement of total bisphenol A in human urine was developed and validated. The method utilized liquid/liquid extraction with 1-chlorobutane and a human urine aliquot size of 800 mu L. Chromatography was performed on an Acquity UPLC (R) system with a Kinetex (R) Phenyl-Hexyl column. Mass spectrometric analysis was with negative electrospray ionization on a Quattro Premier XE (TM). The surrogate matrix method was used for the preparation of calibration standards in synthetic urine due to the presence of BPA in control human urine. The validated calibration range was 0.75-20 ng/mL with a limit of detection of 0.1 ng/mL. The internal standard was d(16)-bisphenol A. Method validation utilized quality control samples at three concentrations in both synthetic urine and human urine. Bisphenol A mono-glucuronide was fortified in synthetic urine in each analytical run to monitor the enzymatic conversion of the glucuronide conjugate to BPA by beta-glucuronidase. Validated method parameters included linearity, accuracy, precision, integrity of dilution, selectivity, re-injection reproducibility, recovery/matrix effect, solution stability, and matrix stability in human urine. Acceptance criteria for analytical standards and QCs were +/-20% of nominal concentration. Matrix stability in human urine was validated after 24 h at ambient temperature, after three freeze/thaw cycles, and after frozen storage at -20 degrees C and -80 degrees C for up to 218 days. The method has been applied to the analysis of over 1750 human urine samples from a biomonitoring study. The median and mean urine BPA concentrations were 2.71 ng/mL and 4.75 ng/mL, respectively. (C) 2014 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据