期刊
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES
卷 944, 期 -, 页码 152-156出版社
ELSEVIER SCIENCE BV
DOI: 10.1016/j.jchromb.2013.11.009
关键词
Bisphenol A; Bisphenol F; Bisphenol S; Exposure; HPLC-MS/MS; Urine
资金
- U.S. Department of Energy
- CDC
Human exposure to bisphenol A (BPA) is widespread. However, in recent years, bisphenol analogs such as bisphenol S (BPS) and bisphenol F (BPF) are replacing BPA in the production of some consumer products. Because human exposure to these alternative bisphenols may occur, biomonitoring of these bisphenol analogs is warranted. In the present study, we developed and validated a sensitive and selective method that uses on-line solid phase extraction coupled to high performance liquid chromatography-isotope dilution tandem mass spectrometry with peak focusing to measure BPA, BPF, BPS, and 11 other environmental phenols in urine. The method required a small amount of sample (100 ILL) and minimal sample pretreatment. The limits of detection were 0.03 ng/mL (BPS), 0.06 ng/mL (BPF), 0.10 ng/mL (BPA), and ranged from 0.1 ng/mL to 1.0 ng/mL for the other 11 phenols. In 100 urine samples collected in 2009-2012 from a convenience group of anonymous adults in the United States, of the three bisphenols, we detected BPA at the highest frequency and median concentrations (95%, 0.72 ng/mL), followed by BPS (78%, 0.13 ng/mL) and BPF (55%, 0.08 ng/mL). This sensitive, rugged, and labor and cost-effective method could be used for the analysis of large number of samples for epidemiologic studies. Published by Elsevier B.V.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐