Development and validation of a rapid capillary zone electrophoresis method for determining charge variants of mAb

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.jchromb.2012.08.022

关键词

Capillary zone electrophoresis; Therapeutic monoclonal antibody; Optimization; Validation; Application

资金

  1. National Natural Science Foundation of China [31100264]

向作者/读者索取更多资源

This work aimed to develop a rapid capillary zone electrophoresis (CZE) method to provide abundant purity and identity information of monoclonal antibodies. The CZE running buffer system was optimized to be 20 mM acetate-acetic acid (pH 6.0) together with the co-addition of 0.3% polyethylene oxide (PEO) and 2 mM triethylenetetramine (TETA), which was further tested with advantages on the peak resolution improvements. The conditioning period was scheduled to 1 min for both 0.1 M HCl and CZE running buffer to reduce total separation time. Additionally, the applied voltage and effective separation length were optimized at 30 kV and 20 cm separately. Compared with the method reported by Yan [1], this newly developed method showed a higher resolution in separating the two unknown basic peaks by testing monoclonal antibody sample (mAb1). The further validation results showed that for all five of charge isoform peaks of test mAb1, repeatability, intraday and interday precision had a RSD less than 0.58% for migration time and less than 3.18% for corrected area percent. The correlation coefficients of more than 0.98 for all peaks also demonstrated the good linearity for the method. In addition to the application of distinguishing intact antibody from C-terminal Lys variants, the method also has advantage in separating the Fab, Fe and intact antibody-relevant substances quickly, which facilitated the rough evaluation of papain induced digestion. (C) 2012 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据