Measurement of 25-OH-vitamin D in human serum using liquid chromatography tandem-mass spectrometry with comparison to radioimmunoassay and automated immunoassay

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.jchromb.2010.03.035

关键词

25-Hydroxyvitamin D; LC-MS/MS; Radioimmunoassay; Method comparison

向作者/读者索取更多资源

The plasma 25-OH vitamin D concentration is a reliable biomarker for vitamin D status but assay's variability makes adequate monitoring of vitamin D status difficult. We employed isotope-dilution liquid chromatography (LC) tandem-mass spectrometry (MS/MS) for the measurement of both 25-OH vitamin D(3) and 25-OH vitamin D(2) in human serum. Hexadeuterium labelled 25-OH vitamin D(3) internal standard (IS) was added to calibrators (prepared in phosphate-buffered saline with 60 g/L albumin), controls or patient sera and 25-OH vitamin D metabolites were released from vitamin D binding protein by adding sodium hydroxide prior to protein precipitation by acetonitrile/methanol (9:1, v/v). Subsequent off-line solid-phase extraction was followed by chromatographic separation on a C-18 column using a water/methanol/ammonium acetate gradient. Detection was by Atmospheric Pressure Electrospray Ionisation (AP-EI) followed by selected reaction monitoring. We compared the LC-MS/MS assay to the DiaSorin radioimmunoassay (RIA) and a recently re-standardised version of an automated electrochemi-luminescent immunoassay (ECLIA) from Roche Diagnostics. We also analysed external quality control samples from the International Vitamin D External Quality Assessment Scheme (DEQAS) for comparison with other participating laboratories using LC-MS. The method was linear from 5 to at least 550 nmol/L with intra- and interday CV's <= 6% for both 25-OH vitamin D(3) and 25-OH vitamin D(2). Recoveries ranged between 94.9 and 106.9% for 25-OH vitamin D(3) and 82.7 and 100.3% for 25-OH vitamin D(2). Our results for the DEQAS serum pools averaged 7.2% from the overall LC-MS method mean. The DiaSorin RIA agreed well with the LC-MS/MS method (r(2) = 0.90; average bias 1.61 nmol/L), the Roche ECLIA considerably disagreed (r2 = 0.58: bias 10.13 nmol/L). This LC-MS/MS method is reliable and robust for the measurement of both 25-OH vitamin D(3) and 25-OH vitamin D(2) in human serum. (C) 2010 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据