4.5 Article

Validated GC/MS method for the simultaneous determination of clozapine and norclozapine in human plasma. Application in psychiatric patients under clozapine treatment

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.jchromb.2010.07.001

关键词

Clozapine; Norclozapine; GC/MS; Plasma; Patients

向作者/读者索取更多资源

A sensitive and specific GC/MS method for the determination of clozapine (CLZ) and its major metabolite norclozapine (NCLZ), in plasma has been developed, optimized and validated. Specimen preparation includes solid-phase extraction of both analytes using Bond-Elut Certify cartridge and further derivatization with TFAA. Clozapine-d8 was used as internal standard for the determination of CLZ and NCLZ. Limits of detection were 0.45 ng/mL for CLZ and 1.59 ng/mL for NCLZ, while limits of quantification were 1.37 ng/mL for CLZ and 4.8 ng/mL for NCLZ, as calculated by the calibration curves. The calibration curves were linear up to 600 ng/mL for CLZ and NCLZ. Absolute recovery ranged from 82.22% to 95.35% for both analytes. Intra- and interday accuracy was less than 7.13% and -12.52%, respectively, while intra- and interday precision was between 9.47% and 12.07%. respectively, for CLZ and NCLZ. The method covers all therapeutic range and proved suitable for the determination of CLZ and NCLZ not only in psychiatric patients but also in forensic cases with clozapine implication. (C) 2010 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据