4.5 Article

Liquid chromatography atmospheric pressure chemical ionization tandem mass spectrometry method for the quantification of pregabalin in human plasma

出版社

ELSEVIER
DOI: 10.1016/j.jchromb.2009.10.004

关键词

Pregabalin; Liquid chromatography-tandem mass spectrometry; Atmospheric pressure chemical ionization; Human plasma; Pharmacokinetic study

资金

  1. Suven Life Sciences Ltd, Hyderabad

向作者/读者索取更多资源

A sensitive high-performance liquid chromatography positive ion atmospheric pressure chemical ionization tandem mass spectrometry method was developed and validated for the quantification of pregabalin in human plasma. Following liquid-liquid extraction, the analyte was separated using an isocratic mobile phase on a reverse-phase column and analyzed by MS/MS in the multiple reaction monitoring mode using the respective [M+H](+) ions, m/z 160-142 for pregabalin and m/z 482-258 for the internal standard. The assay exhibited a linear dynamic range of 1-10,000 ng/mL for pregabalin in human plasma. The lower limit of quantification was 1 ng/mL with a relative standard deviation of less than 11.4%. Acceptable precision and accuracy were obtained for concentrations over the standard curve range. A run time of 4.0 min for each sample made it possible to analyze more than 300 human plasma samples per day. The validated method has been successfully used to analyze human plasma samples for application in pharmacokinetic studies. (C) 2009 Elsevier B.V. All rights reserved.