4.5 Article

Validation of a gas chromatography-mass spectrometry method for the analysis of sterol oxidation products in serum

出版社

ELSEVIER
DOI: 10.1016/j.jchromb.2008.01.036

关键词

serum; sterols; COPs; SOPs; linearity; recovery; repeatability

向作者/读者索取更多资源

A validated gas chromatography-mass spectrometry (GC-MS) detection method for the quantitative analysis of sterol oxidation products (SOPs) in serum is described. After a lipid extraction procedure with chloroform-methanol, a cold saponification and purification by solid phase extraction, oxysterols were derivatizated to form trimethyl-sylil-ethers which were subjected to GC-MS analysis. Calibration curves for cholesterol oxidation products showed determination coefficient (R-2) of 1.0, with low values for the coefficient of variation of the response factors (<1%). Detection and quantification limits were below 5 ng/mL and 10 ng/mL, respectively. Recovery data were between 77.65% and 110.29% (CV < 10% for all compounds). Good results were obtained for within- and between-day repeatability, with values below 10%. In conclusion, the method performed is suitable for the determination and quantification of SOPs in serum. (C) 2008 Elsevier B.V. All rights reserved.