期刊
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES
卷 867, 期 2, 页码 213-218出版社
ELSEVIER SCIENCE BV
DOI: 10.1016/j.jchromb.2008.04.019
关键词
HPLC-ECD; LC-MS/MS; artesunate; dihydroartemisinin; validation
High-performance liquid chromatography with reductive electrochemical detection (HPLC-ECD) method has been used for assaying artemisinins since 1985. Although the methods have been remarkably improved, tandem mass spectrometry (LC-MS/MS) systems with significant advantages have gradually replaced HPLC-ECD to analyze artesunate and dihydroartemisinin in plasma. In the present study, the two methods were evaluated for linearity, quantitation limits, selectivity, precision, and accuracy. The HPLC-ECD performed well in terms of various validation parameters, and showed a good agreement with the LC-MS/MS when calibrated in plasma. However, the major benefit of LC-MS/MS is that it requires only one-tenth the plasma volume needed by HPLC-ECD assay. Published by Elsevier B.V.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。