Detailed characterization of the kinetic performance of first and second generation silica monolithic columns for reversed-phase chromatography separations

期刊

JOURNAL OF CHROMATOGRAPHY A
卷 1325, 期 -, 页码 72-82

出版社

ELSEVIER
DOI: 10.1016/j.chroma.2013.11.047

关键词

Second generation silica monoliths; Kinetic performance; Characteristic length; Flow resistance

向作者/读者索取更多资源

The kinetic performance of commercially available first generation and prototype second generation silica monoliths has been investigated for 2.0 mm and 3.0-3.2 mm inner diameter columns. It is demonstrated that the altered sol-gel process employed for the production of second generation monoliths results in structures with a smaller characteristic size leading to an improved peak shape and higher efficiencies. The permeability of the columns however, decreases significantly due to the smaller throughpore and skeleton sizes. Scanning electron microscopy pictures suggest the first generation monoliths have cylindrical skeleton branches, whereas the second generation monoliths rather have skeleton branches that resemble a single chain of spherical globules. Using recently established correlations for the flow resistance of cylindrical and globule chain type monolithic structures, it is demonstrated that the higher flow resistance of the second generation monoliths can be entirely attributed to their smaller skeleton sizes, which is also evident from the external porosity that is largely the same for both monolith generations (epsilon(e) similar to 0.65). The recorded van Deemter plots show a clear improvement in efficiency for the second generation monoliths (minimal plate heights of 13.6-14.1 mu m for the first and 6.5-8.2 mu m for the second generation, when assessing the plate count using the Foley-Dorsey method). The corresponding kinetic plots, however, indicate that the much reduced permeability of the second generation monoliths results in kinetic performances (time needed to achieve a given efficiency) which are only better than those of the first generation for plate counts up to N similar to 45,000. For more complex samples (N >= 50,000), the first generation monoliths can intrinsically still provide faster analysis due to their high permeability. It is also demonstrated that - despite the improved efficiency of the second generation monoliths in the practical range of separations (N = 10,000-50,000) - these columns can still not compete with state-of-the-art core-shell particle columns when all columns are evaluated at their own maximum operating pressure (200 bar for the monolithic columns, 600 bar for core-shell columns). It is suggested that monolithic columns will only become competitive with these high efficiency particle columns when further improvements to their production process are made and their pressure resistance is raised. (C) 2013 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据