期刊
JOURNAL OF CHROMATOGRAPHY A
卷 1349, 期 -, 页码 69-79出版社
ELSEVIER
DOI: 10.1016/j.chroma.2014.04.100
关键词
Stir-bar sorptive extraction; GC-MS/MS; UHPLC-MS/MS; Endocrine disrupting chemicals; Human breast milk
In the present work, two specific, accurate and sensitive methods for the determination of endocrine disrupting chemicals (EDCs) in human breast milk are developed and validated. Bisphenol A and its main chlorinated derivatives, five benzophenone-UV filters and four parabens were selected as target analytes. The method involves a stir-bar sorptive extraction (SBSE) procedure followed by a solvent desorption prior to GC-MS/MS or UHPLC-MS/MS analysis. A derivatization step is also necessary when GC analysis is performed. The GC column used was a capillary HP-5MS with a run time of 26 min. For UHPLC analysis, the stationary phase was a non-polar Acquity UPLC (R) BEH C18 column and the run time was 10 min. In both cases, the analytes were detected and quantified using a triple quadrupole mass spectrometer (QqQ). Quality parameters such as linearity, accuracy (trueness and precision), sensitivity and selectivity were examined and yielded good results. The limits of quantification (LOQs) ranged from 0.3 to 5.0 ng mL(-1) for GC and from 0.2 to 1.0 ng mL(-1) for LC. The relative standard deviation (RSD) was lower than 15% and the recoveries ranged from 92 to 114% in all cases, being slightly unfavorable the results obtained with LC. The methods were satisfactorily applied for the determination of target compounds in human milk samples from 10 randomly selected women. (C) 2014 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据