4.6 Article

Chemical anchoring of lauryl methacrylate-based reversed phase monolith to 1/16 o.d. polyetheretherketone tubing

期刊

JOURNAL OF CHROMATOGRAPHY A
卷 1242, 期 -, 页码 59-66

出版社

ELSEVIER
DOI: 10.1016/j.chroma.2012.04.030

关键词

Reversed-phase HPLC; PEEK; Organic polymer monolith; Microbore column; Lauryl-methacrylate

资金

  1. New Energy and Industrial Technology Development Organization (NEDO) of Japan [07C46215a]
  2. Japan Society for the Promotion of Science [21350041]
  3. Grants-in-Aid for Scientific Research [21350041] Funding Source: KAKEN

向作者/读者索取更多资源

In this paper, we describe a method for the preparation of easy-to-use reversed-phase monolithic microbore columns. Polyetheretherketone (PEEK) tubing with an outer diameter of 1/16 '' and an inner diameter of 1.0 mm was used as a column housing (empty column), and in it lauryl methacrylate (LMA) was copolymerized with ethylene dimethacrylate (EDMA). In order to chemically anchor the polymer monolith to the tube wall, the inner wall surface was pretreated by the following two-step procedure. (1) 50% sulfuric acid was filled into the PEEK tubing and left to stand for 6 h to generate sulfonate groups on the surface. (2) After washing with Milli-Q water, the sulfonated PEEK surface was brought into contact with 1 M glycidyl methacrylate in dichloromethane (or acetone) at 40 degrees C for 4 h to introduce methacryloyl groups via the reaction of sulfonate groups and epoxy groups. Mechanical strength and column efficiency of the resulting monoliths were evaluated through the separation of a series of alkylbenzenes in acetonitrile-water (50:50, v/v) eluent over the flow rate range of 50-750 mu L/min (corresponding to 1.7-25.5 mm/s). The poly(LMA-co-EDMA) monolith provided acceptable column efficiency of 2000 theoretical plates/10 cm (HETP value of 50 mu m) for amylbenzene (separation factor k = 40) and low flow resistance of 0.5 MPa/10 cm at a normal flow rate of 50 mu L/min. The methacryloylated PEEK tubing tightly held the monolith, and the monolithic column exhibited good pressure resistance up to 15 MPa. allowing rapid separation at a 15-20 fold higher flow rate than normal. (C) 2012 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据