4.6 Article

On the extra-column band-broadening contributions of modern, very high pressure liquid chromatographs using 2 1 mm I D columns packed with sub-2 μm particles

期刊

JOURNAL OF CHROMATOGRAPHY A
卷 1217, 期 49, 页码 7677-7689

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.chroma.2010.10.016

关键词

Instrumentations; Band broadening; 1290 Infinity HPLC System; Acquity; Extra column effects; Efficiency; BEH C-18

资金

  1. National Science Foundation [CHE-06-08659]
  2. University of Tennessee
  3. Oak Ridge National Laboratory

向作者/读者索取更多资源

The efficiencies of two narrow bore columns (100 mm and 50 mm x 2 1 mm) packed with 1 7 m totally porous BEH-C-18 particles were measured on two very high pressure liquid chromatographs (Acquity from Waters and 1290 Infinity H PLC System from Agilent) operating at maximum pressures of 1034 and 1200 bar respectively The probe compounds were a mixture of uracil acetophenone toluene and naphthalene eluted in a 50/50 (v/v) solution of acetonitrile and water at 303 K with a flow rate of 0 40 mL/min The apparent efficiencies of columns which lumps the consequences of band broadening due to the column and the system contributions may depend much on the extra-column volumes of the instruments used Actually it is known for a long time that the apparent column performance is strongly affected by the instrument characteristics including the diameter of the connecting tubes the injection technique (with or without needle seat capillary) and the detection cell volume When the 1290 Infinity HPLC System is equipped with a needle seat an Inlet and an outlet connecting capillary tube with inner diameters around 115 mu m its extra-column variance for a 0 1 mu l injection volume is 9 2 mu L-2 while that of the Acquity instrument is 6 9 mu L-2 Minor modifications suggested by their respective manufacturers allowed significant reductions of these variances to 6 2 and 3 9 mu(2) respectively Yet in their optimized configurations and for weakly retained compounds (k similar or equal to 1) these modern sophisticated instruments cannot provide more than 75% (1290 Infinity) and 85% (Acquity) of the maximum efficiency of a 2 1 mm x 50 mm BEH column For more strongly retained compounds (k>4) in contrast they are both able to provide more than 95% of the maximum expected efficiency (C) 2010 Elsevier B V All rights reserved

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据