4.6 Article

Evaluation of 1.0 mm i.d. column performances on ultra high pressure liquid chromatography instrumentation

期刊

JOURNAL OF CHROMATOGRAPHY A
卷 1217, 期 30, 页码 4925-4933

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.chroma.2010.05.044

关键词

UHPLC; 1.0 mm column; Gradient space; Peak capacity

向作者/读者索取更多资源

The present study focuses on the evaluation of 1.0 mm i.d. (internal diameter) columns on a commercial Ultra-High Pressure system. These systems have been developed specifically to operate columns with small volumes, typically 2.1 mm id., by reducing extra-column volume dispersion. The use of columns with smaller id. results in a reduced solvent consumption and required sample volume. The evaluation of the columns was carried out with samples containing neutral and pharmaceutical compounds. In isocratic mode, the extra-column volume produced additional band broadening leading to poor performances compared to equivalent 2.1 mm i.d. columns. By increasing the length of the column, the influence of the extra-column bandpreading could be reduced and 75,000 plates were obtained when four columns were coupled. In gradient mode, the effect of the extra-column contribution on efficiency was limited and about 80% of the performance of the 2.1 mm i.d. columns was obtained. Optimum conditions in gradient mode were further investigated by changing flow rate, gradient time and column length. A different approach of the calculation of peak capacity was also considered for the comparison of the influence of these different parameters. (C) 2010 Elsevier B.V. All rights reserved.