期刊
JOURNAL OF CHROMATOGRAPHY A
卷 1217, 期 19, 页码 3193-3200出版社
ELSEVIER
DOI: 10.1016/j.chroma.2010.02.001
关键词
Robustness of HPLC methods; Method development; Optimization; Quality by Design; Design Space; 3D-Computer modelling software; Validation; QbD; ICH Q8
The present paper describes a multifactorial optimization of 4 critical HPLC method parameters, i.e. gradient time (t(G)), temperature (T), pH and ternary composition (B-1:B-2) based on 36 experiments. The effect of these experimental variables on critical resolution and selectivity was carried out in such a way as to systematically vary all four factors simultaneously. The basic element is a gradient time-temperature (t(G)-T) plane, which is repeated at three different pH's of the eluent A and at three different ternary compositions of eluent B between methanol and acetonitrile. The so-defined volume enables the investigation of the critical resolution for a part of the Design Space of a given sample. Further improvement of the analysis time, with conservation of the previously optimized selectivity, was possible by reducing the gradient time and increasing the flow rate. Multidimensional robust regions were successfully defined and graphically depicted. (C) 2010 Elsevier B.V. All rights reserved.