4.6 Article

Dispersive liquid-liquid microextraction combined with semi-automated in-syringe back extraction as a new approach for the sample preparation of ionizable organic compounds prior to liquid chromatography

期刊

JOURNAL OF CHROMATOGRAPHY A
卷 1198, 期 -, 页码 1-6

出版社

ELSEVIER
DOI: 10.1016/j.chroma.2008.05.007

关键词

dispersive liquid-liquid microextraction; in-syringe back extraction; ionizable organic compounds; liquid chromatography

向作者/读者索取更多资源

Dispersive liquid-liquid microextraction (DLLME) followed by a newly designed semi-automated in-syringe back extraction technique has been developed as an extraction methodology for the extraction of polar organic compounds prior to liquid chromatography (LC) measurement. The method is based on the formation of tiny droplets of the extractant in the sample solution using water-immiscible organic solvent (extractant) dissolved in a water-miscible organic dispersive solvent. Extraction of the analytes from aqueous sample into the dispersed organic droplets took place. The extracting organic phase was separated by centrifuging and the sedimented phase was withdrawn into a syringe. Then in-syringe back extraction was utilized to extract the analytes into an aqueous solution prior to LC analysis. Clenbuterol (CB), a basic organic compound used as a model, was extracted from a basified aqueous sample using 25 mu L tetrachloroethylene (TCE, extraction solvent) dissolved in 500 mu L acetone (as a dispersive solvent). After separation of the organic extracting phase by centrifuging, CB enriched in TCE phase was back extracted into 10 mu L of 1% aqueous formic acid (FA) within the syringe. Back extraction was facilitated by repeatedly moving the plunger back and forth within the barrel of syringe, assisted by a syringe pump. Due to the plunger movement, a thin organic film is formed on the inner layer of the syringe that comes in contact with the acidic aqueous phase. Here, CB, a basic analyte, will be protonated and back extracted into FA. Various parameters affecting the extraction efficiency, viz., choice of extraction and dispersive solvent, salt effect, speed of syringe pump, back extraction time period, effect of concentration of base and acid, were evaluated. Under optimum conditions, precision, linearity (correlation coefficient, r(2) = 0.9966 over the concentration range of 10-1000 ng mL(-1) CB), detection limit (4.9 ng mL(-1)), enrichment factor (175), relative recovery (97%) had been obtained. The applicability of this newly developed method was investigated for the analysis of CB in the water samples from river, lake and stream water. (c) 2008 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据