4.6 Article

Creating a fatty acid methyl ester database for lipid profiling in a single drop of human blood using high resolution capillary gas chromatography and mass spectrometry

期刊

JOURNAL OF CHROMATOGRAPHY A
卷 1211, 期 1-2, 页码 120-128

出版社

ELSEVIER
DOI: 10.1016/j.chroma.2008.09.066

关键词

Capillary gas chromatography; Mass spectrometry; Fatty acid methyl esters; Blood; Retention time locking; Database

向作者/读者索取更多资源

Capillary gas chromatography (CGC) in combination with mass spectrometry (MS) was optimized for the separation and detection of the fatty acids occurring in the lipid fraction of blood. A fingertip blood sample (ca. 50 mu L) was transesterified into the methyl esters and analyzed on a 100 m x 0.25 mm ID column coated with a biscyanopropyl polysiloxane (HP-88) stationary phase. The method was retention time locked. Programmed temperature vaporization injection (PTV) in the solvent venting mode was applied to minimize the sample size, while maintaining high sensitivity. The total analysis time was ca. 60 min. Retention times and both electron impact (EI) and positive chemical ionization (PCI) mass spectrometry were combined to elucidate the fatty acids according to alkyl chain, degree of unsaturation and position of the double bonds. Using extracted ion chromatograms about 100 fatty acids and related compounds Were detected in blood samples and most of them were identified. This work resulted in a very large fatty acid methyl esters database, containing retention time and mass spectral information that will be applied to metabolomic studies. (C) 2008 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据