期刊
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME
卷 93B, 期 5, 页码 616-621出版社
BRITISH EDITORIAL SOC BONE JOINT SURGERY
DOI: 10.1302/0301-620X.93B5.25539
关键词
-
We biomechanically investigated whether the standard dynamic hip screw (DHS) or the DHS blade achieves better fixation in bone with regard to resistance to pushout, pullout and torsional stability. The experiments were undertaken in an artificial bone substrate in the form of polyurethane foam blocks with predefined mechanical properties. Pushout tests were also repeated in cadaveric femoral heads. The results showed that the DHS blade outperformed the DHS with regard to the two most important characteristics of implant fixation, namely resistance to pushout and rotational stability. We concluded that the DHS blade was the superior implant in this study.