The Metasul metal-on-metal articulation in primary total hip replacement: Clinical and radiological results at ten years

期刊

JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME
卷 90B, 期 10, 页码 1278-1283

出版社

BRITISH EDITORIAL SOC BONE JOINT SURGERY
DOI: 10.1302/0301-620X.90B10.20378

关键词

-

向作者/读者索取更多资源

We report the clinical and radiological outcome at ten years of 104 primary total hip replacements (100 patients) using the Metasul metal-on-metal bearing. Of these, 52 had a cemented Stuehmer-Weber polyethylene acetabular component with a Metasul bearing and 52 had an uncemented Allofit acetabular component with a Metasul liner. A total of 15 patients (16 hips) died before their follow-up at ten years and three were lost to follow-up. The study group therefore comprised 82 patients (85 hips). The mean Oxford score at ten years was 20.7 (12 to 42). Six of 85 hips requires revision surgery. One was performed because of infection, one for aseptic loosening of the acetabular component and four because of unexplained pain. Histological examination showed and aseptic lymphocytic vasculitis associated lesion-type tissue response in two of these. Continued follow-up is advocated in order to monitor the long-term performance of the Metasul bearing and tissue responses to metal debris.