The Seven-Year Wear of Highly Cross-Linked Polyethylene in Total Hip Arthroplasty A Double-Blind, Randomized Controlled Trial Using Radiostereometric Analysis

期刊

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.2106/JBJS.J.00287

关键词

-

资金

  1. Zimmer, Inc.
  2. Biomet
  3. Corin
  4. Wright Medical
  5. Zimmer

向作者/读者索取更多资源

Background: The use of highly cross-linked polyethylene is now commonplace in total hip arthroplasty. Hip simulator studies and short-term in vivo measurements have suggested that the Wear rate of highly cross-linked polyethylene is significantly less than that of conventional ultra-high molecular weight polyethylene. However, long-term data to support its use are limited. The aim of this study was to compare the intermediate-term steady-state wear of highly cross-linked polyethylene compared with that of conventional ultra-high molecular weight polyethylene acetabular liners in a prospective, double-blind, randomized controlled trial with use of radiostereometric analysis. Methods: Fifty-four patients were randomized to receive hip replacements with either conventional ultra-high molecular weight polyethylene acetabular liners (Zimmer) or highly cross-linked polyethylene liners (Longevity; Zimmer). All patients received a cemented, collarless, polished, tapered femoral component (CPT; Zimmer) and an uncemented acetabular component (Trilogy; Zimmer). Clinical outcomes were assessed and the three-dimensional penetration of the head into the socket was determined for a minimum of seven years. Linear regression was used to calculate the steady-state wear rate following the creep-dominated penetration seen during the first year. Results: At a minimum of seven years postoperatively, the mean total femoral head penetration was significantly lower in the highly cross-linked polyethylene group (0.33 mm; 95% confidence interval [Cl], +/- 0.10 mm) than it was in the ultra-high molecular weight polyethylene group (0.55 mm; 95% Cl, +/- 0.10 mm) (p = 0.005). The mean steady-state wear rate of highly cross-linked polyethylene was 0.005 mm/yr (95% Cl, +/- 0.015 mm/yr), compared with 0.037 mm/yr (95% Cl, +/- 0.019 mm/yr) for conventional ultra-high molecular weight polyethylene (p = 0.007). No patient in the highly cross-linked polyethylene group had a wear rate above the osteolysis threshold of 0.1 mm/yr, compared with 9% of patients in the ultra-high molecular weight polyethylene group. Conclusions: This study demonstrates that highly cross-linked polyethylene has a significantly lower steady-state wear rate compared with that of conventional ultra-high molecular weight polyethylene. Longer-term follow-up is required to determine if this will translate into improved clinical performance and longevity of these implants.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据