4.5 Article

Raman spectroscopy for clinical-level detection of heparin in serum by partial least-squares analysis

期刊

JOURNAL OF BIOMEDICAL OPTICS
卷 18, 期 2, 页码 -

出版社

SPIE-SOC PHOTO-OPTICAL INSTRUMENTATION ENGINEERS
DOI: 10.1117/1.JBO.18.2.027010

关键词

Raman spectroscopy; heparin; serum; partial least squares

资金

  1. Ontario Centres of Excellence
  2. Technology Transfer and Business Enterprise at the University of Ottawa
  3. Animal Care and Veterinary Service

向作者/读者索取更多资源

Heparin is the most widely used anti-coagulant for the prevention of blood clots in patients undergoing certain types of surgeries including open heart surgeries and dialysis. The precise monitoring of heparin amount in patients' blood is crucial for reducing the morbidity and mortality in surgical environments. Based upon these considerations, we have used Raman spectroscopy in conjunction with partial least squares (PLS) analysis to measure heparin concentration at clinical level which is less than 10 United States Pharmacopeia (USP) in serum. The PLS calibration model was constructed from the Raman spectra of different concentrations of heparin in serum. It showed a high coefficient of determination (R-2 > 0.91) between the spectral data and heparin level in serum along with a low root mean square error of prediction similar to 4 USP/ml. It enabled the detection of extremely low concentrations of heparin in serum (similar to 8 USP/ml) as desirable in clinical environment. The proposed optical method has the potential of being implemented as the point-of-care testing procedure during surgeries, where the interest is to rapidly monitor low concentrations of heparin in patient's blood. (C) 2013 Society of Photo-Optical Instrumentation Engineers (SPIE) [DOI: 10.1117/1.JBO.18.2.027010]

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据