Regional tissue oxygen saturation: comparability and reproducibility of different devices

期刊

JOURNAL OF BIOMEDICAL OPTICS
卷 16, 期 5, 页码 -

出版社

SPIE-SOC PHOTOPTICAL INSTRUMENTATION ENGINEERS
DOI: 10.1117/1.3575647

关键词

tissue oxygenation index; rSO2; NIRO300; INVOS5100; comparability; reproducibility

向作者/读者索取更多资源

Comparability and reproducibility of different near-infrared spectroscopy devices measuring regional tissue oxygen saturation remain poor. Aim of the present study was to compare values and reproducibility of cerebral/peripheral tissue-oxygenation-index (TOI; NIRO 300, Hamamatsu (R), Japan) with cerebral/peripheral regional-oxygen-saturation (rSO; INVOS5100, Somanetics (R), USA), and to analyze the influence of quality criteria. Methods: cTOI and crSO2 were measured on the left forehead, pTOI and prSO2 were measured on the left calf. To analyse reproducibility, optodes were reapplied five times. A quality criterion was introduced for cTOI, crSO2 and prSO2. For pTOI quality criteria were introduced in combination with a venous occlusion technique. Results: Cerebral measurements were performed in 37 neonates. cTOI (72.7+/-6.2%) was lower than crSO2 (83.3+/-5.8%) (p < 0.001). The mean difference between cTOI and crSO2 was 10%. Mean standard deviations of cTOI and crSO2 were similar (cTOI: 4.9+/-3.6; crSO2: 4.5+/-2.6). Peripheral measurements were performed in 39 neonates. pTOI (66.0+/-7.9%) was lower than prSO2 (82.0+/-7.0%)(p < 0.001). The mean difference between pTOI and prSO2 was 15%. Mean standard deviations of pTOI (3.7+/-2.6%) were lower than of prSO2 (5.0+/-3.0%) (p = 0.047). Conclusion: TOI values were significantly lower than rSO2 values, in cerebral and peripheral measurements. Reproducibility was higher for pTOI than for prSO2. (C) 2011 Society of Photo-Optical Instrumentation Engineers (SPIE). [DOI: 10.1117/1.3575647]

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据