4.5 Article

Multipoint fluorescence correlation spectroscopy with total internal reflection fluorescence microscope

期刊

JOURNAL OF BIOMEDICAL OPTICS
卷 14, 期 1, 页码 -

出版社

SPIE-SOC PHOTO-OPTICAL INSTRUMENTATION ENGINEERS
DOI: 10.1117/1.3080723

关键词

fluorescence spectroscopy; correlation; total internal reflection fluorescence microscope; cell membranes; molecular dynamics

资金

  1. Japan Society for the Promotion of Science (JSPS) [15370062, 18207010, 19021001, 19038001]
  2. Grants-in-Aid for Scientific Research [19058001] Funding Source: KAKEN

向作者/读者索取更多资源

We report simultaneous determination of diffusion coefficients at different points of a cell membrane using a multipoint fluorescence correlation spectroscopy (FCS) system. A system carrying seven detection areas in the evanescent field is achieved by using seven optical fibers on the image plane in the detection port of an objective-type total internal reflection FCS (TIR-FCS) system. Fluctuation of fluorescence intensity is monitored and evaluated using seven photomultiplier tubes (PMTs) and a newly constructed multichannel correlator. We demonstrate simultaneous-multipoint FCS, with a 3-mu s time resolution, to investigate heterogeneous structures such as cell membranes and membrane-binding molecular dynamics near glass surfaces in live cells. (C) 2009 Society of Photo-Optical Instrumentation Engineers. [DOI: 10.1117/1.3080723]