4.5 Article

Autofluorescence of epithelial tissue: single-photon versus two-photon excitation

期刊

JOURNAL OF BIOMEDICAL OPTICS
卷 13, 期 5, 页码 -

出版社

SPIE-SOC PHOTO-OPTICAL INSTRUMENTATION ENGINEERS
DOI: 10.1117/1.2975866

关键词

autofluorescence; confocal microscopy; two-photon spectroscopy; epithelial tissue

资金

  1. Hong Kong Research Grants Council [HKUST6408/05M]
  2. HKUST [RPC06/07.EG06]

向作者/读者索取更多资源

We instrumented a combined fluorescence spectroscopy and imaging system to characterize the single- and two-photon excited autofluorescence in epithelial tissue. Single-photon fluorescence (SPF) are compared with two-photon fluorescence (TPF) measured at the same location in epithelial tissue. It was found that the SPF and TPF signals excited at corresponding wavelengths are similar in non-keratinized epithelium, but the SPF and TPF spectra in the keratinized epithelium and the stromal layer are significant different. Specifically, the comparison of SPF signals with TPF signals in keratinized epithelial and stromal layers shows that TPF spectral peaks always have about 15-nm redshift with respect to SPF signals, and the TPF spectra are broader than SPF spectra. The results were generally consistent with the SPF and TPF measurements of pure nicotinamide adenine dinucleotide, flavin adenine dinucleotide, keratin and collagen, the major fluorophores in epithelium and stroma, respectively. The double peak structure of TPF spectra measured from keratinized layer suggests that there may be an unknown fluorophore responsible for the spectral peak in the long wavelength region. Furthermore, the TPF signals excited in a wide range of wavelengths provide accurate information on epithelial structure, which is an important advantage of TPF over SPF spectroscopy in the application for the diagnosis of tissue pathology. (C) 2008 Society of Photo-Optical Instrumentation Engineers. [DOI: 10.1117/1.2975866]

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据