Surface characterization of explanted endovascular stents: Evidence of in vivo corrosion

出版社

WILEY
DOI: 10.1002/jbm.b.31698

关键词

stent corrosion; fretting corrosion; pitting corrosion; overlapping stents; ion release

资金

  1. National Institute of Health [NIBIB-BRP 1 RO1 EB001715-01A2, P30-AR4603]
  2. Cyprus Research Promotion Foundation, Nicosia, Cyprus
  3. University of Alabama at Birmingham, Center of Metabolic Bone Disease-Histomorphometry and Molecular Analysis Core Laboratory

向作者/读者索取更多资源

Limited information exists regarding the in vivo stability of endovascular stents. Nine excised human vascular segments with implanted stents (n = 16) manufactured from stainless steel, nickel-titanium, tantalum, and cobalt-based alloys were analyzed. The stent/tissue components were separated using an established tissue dissolution protocol and control and explanted stents were evaluated by digital optical and electron microscopy. Metallic content in surrounding tissues was measured by mass spectroscopy. Surface alterations, consistent with corrosion mediated by electrochemical and mechanical factors, were observed in 9 of the 16 explanted stents and were absent from control stents. Tissue dissolved from around corroded stents corresponded with a higher metallic content. The effect of these changes in the microtopography of stents on their mechanical properties (fatigue strength and fracture limit) in addition to the potential for released metallic debris contributing to the biological mechanisms of in-stent restenosis supports the need for further investigations. (C) 2010 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater 95B: 225-238, 2010.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据