4.5 Article

Thrombogenicity assessment of Pipeline Flex, Pipeline Shield, and FRED flow diverters in an in vitro human blood physiological flow loop model

期刊

JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A
卷 106, 期 12, 页码 3195-3202

出版社

WILEY
DOI: 10.1002/jbm.a.36514

关键词

flow diversion; intracranial aneurysm; thrombosis; surface modification; phosphorylcholine

向作者/读者索取更多资源

Endovascular treatment of intracranial aneurysms with endoluminal flow diverters (single or multiple) has proven to be clinically safe and effective, but is associated with a risk of thromboembolic complications. Recently, a novel biomimetic surface modification with covalently bound phosphorylcholine (Shield Technology (TM)) has shown to reduce the material thrombogenicity of the Pipeline flow diverter. Thrombogenicity of Pipeline Flex, Pipeline Shield, and Flow Redirection Endoluminal Device (FRED) in the presence of human blood under physiological flow conditions-in addition to relative increase in thrombogenicity with multiple devices-remains unknown and was investigated here. Thrombin generation (mean +/- SD; mu g/mL; thrombin-antithrombin complex or TAT) was measured as FRED (30.3 +/- 2.9), Pipeline (13.9 +/- 4.4), Pipeline Shield (0.4 +/- 0.3), and negative control (no device; 0.1 +/- 0.0). Platelet activation (mean +/- SD; IU/mu L; beta-thromboglobulin or beta TG) was measured as FRED (148 +/- 45), Pipeline (92.8 +/- 41), Pipeline Shield (16.2 +/- 3.5), and negative control (2.70 +/- 0.16). FRED was significantly more thrombogenic than Pipeline and Pipeline Shield (p < 0.05) for TAT. Additionally, Pipeline Shield had significantly lower TAT and beta TG than the other devices tested (p < 0.05) and these were comparable to the negative control (p > 0.05). TAT and beta TG scaled proportionately with multiple Pipeline devices (N = 6) but was unaffected by multiple Pipeline Shield (N = 6) devices-the latter being statistically similar to negative control (p > 0.05). (c) 2018 The Authors. Journal Of Biomedical Materials Research Part A Published By Wiley Periodicals, Inc. J Biomed Mater Res Part A: 106A: 3195-3202, 2018.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据