4.5 Article

Host response to microgel coatings on neural electrodes implanted in the brain

期刊

JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A
卷 102, 期 5, 页码 1486-1499

出版社

WILEY
DOI: 10.1002/jbm.a.34799

关键词

microgel; electrode; inflammation; cell adhesion; implantation; polymer coating

资金

  1. National Institutes of Health [F31NS073358, T32EB006343-01A2]
  2. GAANN Fellowship for Drug Design, Development, and Delivery: US Department of Education [P200A090099]
  3. Georgia Tech/Emory Center for the Engineering of Living Tissues
  4. Atlanta Clinical and Translational Science Institute (from the Clinical and Translational Science Award Program) [UL RR025008]
  5. National ESCA and Surface Analysis Center for Biomedical Problems (NESAC/BIO) [NIH EB-002027]

向作者/读者索取更多资源

The performance of neural electrodes implanted in the brain is often limited by host response in the surrounding brain tissue, including astrocytic scar formation, neuronal cell death, and inflammation around the implant. We applied conformal microgel coatings to silicon neural electrodes and examined host responses to microgel-coated and uncoated electrodes following implantation in the rat brain. In vitro analyses demonstrated significantly reduced astrocyte and microglia adhesion to microgel-coated electrodes compared to uncoated controls. Microgel-coated and uncoated electrodes were implanted in the rat brain cortex and the extent of activated microglia and astrocytes as well as neuron density around the implant were evaluated at 1, 4, and 24 weeks postimplantation. Microgel coatings reduced astrocytic recruitment around the implant at later time points. However, microglial response indicated persistence of inflammation in the area around the electrode. Neuronal density around the implanted electrodes was also lower for both implant groups compared to the uninjured control. These results demonstrate that microgel coatings do not significantly improve host responses to implanted neural electrodes and underscore the need for further improvements in implantable materials. (c) 2013 Wiley Periodicals, Inc. J Biomed Mater Res Part A: 102A: 1486-1499, 2014.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据