4.6 Article

Tibial Plateau Coverage in UKA: A Comparison of Patient Specific and Off-The-Shelf Implants

期刊

JOURNAL OF ARTHROPLASTY
卷 29, 期 9, 页码 1694-1698

出版社

CHURCHILL LIVINGSTONE INC MEDICAL PUBLISHERS
DOI: 10.1016/j.arth.2014.03.026

关键词

unicompartmental knee arthroplasty; patient specific implants; tibial component; overhang; CT-based imaging; ConforMIS

向作者/读者索取更多资源

Poor tibial component fit can lead to issues including pain, loosening and subsidence. Morphometric data, from 30 patients undergoing UKA were utilized; comparing size, match and fit between patient-specific and off-the-shelf implants. CT images were prospectively obtained and implants modeled in CAD, utilizing sizing templates with off-the-shelf and CAD designs with patient-specific implants. Virtual surgery was performed, maximizing tibial plateau coverage while minimizing implant overhang. Each implant evaluated to examine tibial fit. Patient-specific implants provided significantly greater cortical rim surface area coverage versus off-the-shelf implants: 77% v. 43% medially and 60% v. 37% laterally. Significantly less cortical rim overhang and undercoverage were observed with patient-specific implants. Patient-specific implants provide superior cortical bone coverage and fit while minimizing overhang and undercoverage seen in off-the-shelf implants. (C) 2014 Elsevier Inc. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据