期刊
JOURNAL OF ARTHROPLASTY
卷 29, 期 11, 页码 2122-2126出版社
CHURCHILL LIVINGSTONE INC MEDICAL PUBLISHERS
DOI: 10.1016/j.arth.2014.06.023
关键词
hip arthroplasty; conservative stem; metaphyseal stem; short stem; mayo stem; aseptic loosening
类别
The aseptic loosening rate of the Mayo hip prosthesis femoral stem was determined in a 44 patient (52 hips) series, by comparing preoperative and postoperative clinical and radiographic parameters, over aminimal 5-year postoperative period. Loosening was obvious before five years in four hips (7.6%), while the stem was considered stable in the remaining 48 hips over a mean 6.1 year period, therefore accounting for a 92.3% Kaplan-Meier survival rate. We conclude that the aseptic loosening rate is too high for the Mayo hip prosthesis femoral stem, as compared to literature data concerning the non-conservative stems. (C) 2014 Elsevier Inc. All rights reserved.