期刊
JOURNAL OF ARTHROPLASTY
卷 28, 期 2, 页码 227-233出版社
CHURCHILL LIVINGSTONE INC MEDICAL PUBLISHERS
DOI: 10.1016/j.arth.2012.05.019
关键词
overhang; tibial component (tray); UKA; MCL impingement; robot
类别
资金
- Canadian Institutes of Health Research Funding Source: Medline
Excessive tibial component overhang during unicompartmental knee arthroplasty (UKA) may cause medial collateral ligament (MCL) impingement, which, in turn, may lead to medial knee pain [Chau et al. Tibial component overhang 226 following unicompartmental knee replacement does it matter? The Knee. 2009;16(5):310-3]. This study examines MCL loads in 6 human cadaveric knees for different levels of overhang using a robotic testing system. The results indicated no statistically significant difference between the baseline MCL load (no overhang) and the 2-mm overhang (P = .261). However, there were significant differences in MCL load between 2- vs 4-mm (P = .012) and 2- vs 6-mm overhang (P = .022). The loads were almost doubled from 2 to 4 mm of overhang. We conclude that, to minimize pain from excessive MCL loading, surgeons should avoid tibial component overhang greater than 2 mm in unicompartmental knee arthroplasties.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐