4.6 Article

Why Are Total Knees Failing Today? Etiology of Total Knee Revision in 2010 and 2011

期刊

JOURNAL OF ARTHROPLASTY
卷 28, 期 8, 页码 116-119

出版社

CHURCHILL LIVINGSTONE INC MEDICAL PUBLISHERS
DOI: 10.1016/j.arth.2013.04.056

关键词

total knee revision; failure; aseptic loosening; instability

向作者/读者索取更多资源

Revision knee data from six joint arthroplasty centers were compiled for 2010 and 2011 to determine mechanism of failure and time to failure. Aseptic loosening was the predominant mechanism of failure (31.2%), followed by instability (18.7%), infection (16.2%), polyethylene wear (10.0%), arthrofibrosis (6.9%), and malalignment (6.6%). Mean time to failure was 5.9 years (range 10 days to 31 years). 35.3% of all revisions occurred less than 2 years after the index arthroplasty, 60.2% in the first 5 years. In contrast to previous reports, polyethylene wear is not a leading failure mechanism and rarely presents before 15 years. Implant performance is not a predominant factor of knee failure. Early failure mechanisms are primarily surgeon-dependent. (C) 2013 Elsevier Inc. All rights reserved.