4.6 Article

Primary Hip Arthroplasty With 28-mm Metasul Articulation

期刊

JOURNAL OF ARTHROPLASTY
卷 26, 期 4, 页码 662-664

出版社

CHURCHILL LIVINGSTONE INC MEDICAL PUBLISHERS
DOI: 10.1016/j.arth.2010.04.021

关键词

total hip arthroplasty; Metasul; articulation; results

资金

  1. Zimmer
  2. Good Samaritan Hospital (Los Angeles, CA)

向作者/读者索取更多资源

This follow-up study reports on 69 patients at mean 13 years with total hip arthroplasty using 28-mm Metasul (Zimmer, Winterthur, Switzerland) metal-on-metal articulation. These results are not transferable to large diameter head metal-on-metal articulations. Four new revisions, 3 for disassociation of the liner and 1 for mechanical loosening of the acetabulum, occurred since the previous report of mean 7.3 years. The prevalent cause of late revision is disassociation, which suggests a high frictional torque or impingement in these articulation surfaces. No revision was done for osteolysis. Overall, of the original 127 hips, 116 (91%) were known to have maintained their original components.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据