4.6 Article

Corrosion of a Hip Stem With a Modular Neck Taper Junction A Retrieval Study of 16 Cases

期刊

JOURNAL OF ARTHROPLASTY
卷 24, 期 7, 页码 1019-1023

出版社

CHURCHILL LIVINGSTONE INC MEDICAL PUBLISHERS
DOI: 10.1016/j.arth.2008.09.009

关键词

retrieval; corrosion; fretting; hip arthroplasty

向作者/读者索取更多资源

Since the early 1990s, there has been a dramatic increase in modular total hip designs ranging from a stern with a proximal taper and modular head, to a distal stem, double taper proximal neck, and modular head. Clinical advantages of the modular neck include intraoperative adjusment of leg length via the neck-head taper and femoral anteversion via the neck-stern taper. Sixteen cases of a double tapered cone, Margron hip prosthesis, were presented for retrieval analysis. Macroscopic inspection, corrosion testing, light microscopy, and scanning electron microscopy were conducted to elucidate mechanisms of failure. In this regard, 6 neck components showed significant fretting, and crevice corrosion of the neck-stern taper with an average implantation time of 39 months compared with the remaining retrievals, which showed no corrosion with and average time in situ of 2.7 months. This retrieval Study demonstrates that even with a modern taper design and corrosion-resistant materials, increased modularity can lead to fretting and crevice corrosion, metal ion generation, and particulate debris that may contribute to periprosthetic osteolysis and loosening.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据